• -40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen
  • -40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen

-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen

Type: Impact Testing Machine
Maxcapacity: Temperature Is -40 Degree
Accuracy Grade: 1
Load Way: Electronic Load
Loading Method: Static Load
Display: Digital
Customization:
Manufacturer/Factory, Trading Company
Gold Member Since 2016

Suppliers with verified business licenses

Shandong, China
Importers and Exporters
The supplier has import and export rights
High Repeat Buyers Choice
More than 50% of buyers repeatedly choose the supplier
Years of Export Experience
The export experience of the supplier is more than 10 years
Exhibition Experience
The supplier had participated in offline trade shows, you can check the Audit Report for more information
to see all verified strength labels (26)

Basic Info.

Model NO.
DWC-40
Control
Manual Control
Weight
100-500Kg
Power Source
AC220V
Cooling Method
Compressor Cooling
Transport Package
Plywood Exporting Package
Specification
200kg 1.3CMB
Trademark
HST
Origin
Jinan, Shandong, China
HS Code
9031809090
Production Capacity
1000 Sets/Month

Product Description

DWC-40 -40~+30 degree Low Temperature Chamber for Cooling Impact Specimen 

  

1. Application
The machine is designed according to Charpy notch Impact Test Method for Metal Material, and adopts compressor cooling technology. It utilizes the heat balance principle and cycle stirring method to realize the constant temperature cooling to impact specimen with the reliable performance. Digital display, automatic control temperature, and it is the ideal equipment for cooling and keeping temperature to impact specimen.
 
2. Main component:
The machine obtains one compressor, heating unit, electronic control box , cooling tank, cooling medium circulation system, automatic warning device, etc.
 
Standards:ASTM, E23-02a, En10045, ISO83,GB/T229-2009 
 
3. Working principle:
Boot refrigeration switch, compressor begin to work, refrigeration system into formal working state. Refrigeration compressor continuous work, when close to the setting temperature, the heating device which is in the cooling bath began to scale with heat.  To balance redundant cold quantity produced by the refrigeration system, in order to achieve the purpose of constant temperature. Stirring can make cooling medium of the cooling  tank  continuously circulation, make the temperature uniform.
 
4. Features:
4.1Imported compressor, DANFUS valve, imported evaporation-condensation machine;
4.2Controlled by intelligent instrument, digital presentation temperature value, automatic control temperature, automatic time and alarm.
4.3high security, refrigerate fast, large volume
 
5. Main Specifications:
Model DWC-40 DWC-60 DWC-80
 
DWC-100
Temperature range +30~-40° +30~-60° +30~-80° +30~-100°
Temperature Control Accuracy         ±0.5°
 
 
 
 
Temperature dropping speed
 
+30°C~0°C
1.2°C/min
+30°C~0°C
2°C/min
+30°C~0°C
2°C/min
+30°C~0°C
2°C/min
0°c~-20°C
0.8°C/min
0°c~-20°C
1.5°C/min
0°c~-20°C
1.5°C/min
0°c~-20°C
1.5°C/min
-20°C~-40°C
0.5°C/min
-20°C~-40°C
1.0°C/min
-20°C~-60°C
1.0°C/min
-20°C~-40°C
1.0°C/min
-40°C~-60°C
0.7°C/min
-60°C~-80°C
0.7°C/min
-60°C~-100°C
0.7°C/min
Cold Room Volume     150×505×240mm
Dimensions  900×505×850 mm 900×505×850 mm 900×505×850 mm 1500×1200×1200 mm
Box Material Steel Spray
Loading the number of samples >60 >60 >60 >200
Mode of refrigeration Compressor refrigeration
Cooling Medium Ethanol or other unfrozen liquid
Power 7Kw 7Kw 7Kw 7Kw

-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen
 
 -40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen-40~+30 Degree Low Temperature Chamber for Cooling Impact Specimen
 