Overview:
HB-3000E adopts the load cell to replace the traditional weight, this innovative design makes the instrument more accurate.
HB-3000E selection of testing force and dwell time is controlled and monitored on the front panel, it is automatic during the process of loading, dwell and unloading.
HB-3000E uses 20x Brinell measuring microscope to measure the diameter of indentation, then the operator traces the Brinell hardness table to get the testing result.
| Product Name | Electronic Brinell Hardness Tester | |
| Model | Vexus SHB-3000E | |
| Testing Force | 62.5Kgf, 100Kgf, 125Kgf, 187.5Kgf, 250Kgf, | |
| 500Kgf, 750Kgf, 1000Kgf, 1500Kgf, 3000Kgf | ||
| 612.9N, 980N, 1226N, 1839N, 2452N, | ||
| 4900N, 7355N, 9800N, 14700N, 29400N | ||
| Testing Range | (8~650) HBW | |
| Hardness Display | Check the Table of Brinell Hardness Scale | |
| Magnification of Microscope | Analogue 20x Microscope | |
| Max. Height Of Specimen | 225mm | |
| Instrument Throat | 135mm | |
| Power Supply | AC220V/50Hz;110V/60Hz | |
| Dimension(LxWxH) | 893 x720 x747mm | |
| Gross/Net Weight | 160Kg/130Kg | |
| Execution Standard | ISO 6506 , ASTM E10, JIS Z2243 , GB/T 231 | |
| Accuracy of Brinell Hardness Testing | ||
| Hardness Range | Max. Tolerance | Repetition |
| HBW≤ 125 | ≤± 3.5 % | ≤ 3.5 % |
| 125 < HBW≤225 | ≤± 2.5 % | ≤ 3.0 % |
| HBW > 225 | ≤± 2.0 % | ≤ 2.5 % |