| Customization: | Available |
|---|---|
| After-sales Service: | Online Technical Support |
| Type: | Vickers Hardness Tester |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Production Introduction:
Heat treatment, carbonization, hardening layer, surface coating, steel, nonferrous metals, small and thin parts, etc.
Technical Parameters
|
Model |
HV-1000Z |
HVS-1000Z |
HVS-1000ZT |
|
Display model |
Small LCD digital display |
Big LCD digital display |
touch screen Digita Display |
|
Hardness values |
Input D1,D2 |
Direct display |
Direct Display |
|
Exchange Scales
|
HRA,HRB,HRC,HRD,HRF,HV,HK,HBW,HR15N,HR30N,HR45N,HR15T,HR30T,HR45T |
||
|
Test force |
10gf (0.098N),25gf (0.245N),50gf (0.49N),100gf (0.98N), 200gf (1.96N),300gf (2.94N),500gf(4.9N),1kgf (9.8N) |
||
|
Turret Type |
Automatic |
||
|
Loading and unloading mode |
Automatic |
||
|
Dwell Time |
Adjustable 0 to 60 seconds |
||
|
Micrometer eyepiece multiple |
10X(optional 15X) |
||
|
Me measurement resolution |
0.25µm (drum reading) |
||
|
Objective magnification |
10X 40X |
||
|
Total magnification |
100X(objective)400X(measurement) |
||
|
measurement range |
150µm |
||
|
light source |
Halogen lamp (brightness adjustable) |
||
|
Hardness conversation |
Brinell, rockwell, surface rockwell |
||
|
Max. Height of Specimen |
80mm |
||
|
Depth at center line |
95mm |
||
|
XY test table size |
100X100mm Max. movement:25*25mm |
||
|
Min. Measuring Unit |
0.01μm |
||
|
RS232,Built-in print,data statistics |
No |
Yes |
Yes |
Scope of Applocation:
HVS-1000ZT Touch screen automatic turret digital display microvickers hardness tester It is mainly used for the measurement of the micro Vickers hardness of various metals and some non-metal materials. It can test various parts (machined parts, forgings, cast iron, steel castings, etc.), non-ferrous metals, thermal after-treatment parts, and various micro And thin parts, glass, ceramics, agate, gemstones, etc. for hardness measurement, at the same time suitable for carbonization, quenching and hardening layer effective depth, coating, surface coating and welding parts heat-affected part measurement. It can also be used to observe the internal metallographic structure of various metal parts and to collect, display and output the image.
It is especially suitable for measuring the hardness gradient distribution curve of the workpiece and the surface hardness of the gear after heat treatment, and the effective hardening layer depth.
I. Technical Parameter:
|
Model |
HVS-1000ZT |
|
Test force |
10gf (0.098N),25gf (0.245N),50gf (0.49N),100gf (0.98N), 200gf (1.96N),300gf (2.94N),500gf(4.9N),1kgf (9.8N) |
|
Min.test unit |
0.025µm Electronic measuring eyepiece |
|
Exchange Scales
|
HRA,HRB,HRC,HRD,HRF,HV,HK,HBW,HR15N,HR30N,HR45N,HR15T,HR30T,HR45T |
|
Measuring range
|
5~3000HV |
|
Test force application method |
Automatic (loading, loading and unloading) |
|
Test microscope magnification |
100X (Observation) 400X(test) |
|
Test power and load retention time |
1~60s |
|
X-Y Test table |
Size:100*100mm Maximum movement:25*25mm |
|
Data output |
LCD display readout, built-in printer and RS-232 interface |
|
Max Height of Specimen
|
110mm |
|
Distance from the center of the head to the outer wall |
110mm |
|
power supply |
AC220V+5%,50-60Hz |
|
Dimensions
|
320*500*570mm |
|
Weight
|
About 40Kg |
Instruction
Model HTMV-1000D (HST-HVS1000) Digital Micro Vickers hardness tester with large screen LCD display, use the menu interface type structure, can be in operation panel choose hardness HV or rod HK, test of hardness value, automatic input, automatic display. For all kinds of hardness value mutual conversion, and the testing result is automatic storage, processing, printing with RS-232 interface, and computer online. This machine is digital, high degree of automation. Hardness value error can be modified through softwaret, hardness value more accurate meets the requirement.
Technical Parameter:
| MODEL | HTMV-1000D (HST-HVS1000) |
| Test Force | 10gf,25gf,50gf,100gf,200gf,300gf,500gf,1kgf |
| 0.098N,0.245N,0.49N,0.98N,1.96N,2.94N,4.9N,9.8N | |
| Convert Scales | HRA,HRB,HRC,HRD,HRF,HV,HK,HBW,HBS,HR15N,HR30N,HR45N,HR15T,HR30T,HR45T |
| Min. Unit of measurement | 0.0025μm |
| Test Mode | HV/HK |
| Hardness Test Range | 5~3000HV |
| Test Force Application Method | Automatic (loading, dwelling, unloading) |
| Turret Method | Manual |
| Objective Lens | 10X,40X |
| Eyepiece | 10X |
| Total Magnification | 100X (observation) 400X (measurement) |
| Test Force Dwell Time | 1~60s |
| X-Y Table Size | 100*100mm |
| X-Y Test Table Maximum Movement | 25*25mm |
| Data Output | LCD display readout, built-in printer |
| Max. Height of Specimen | 110mm |
| The Distance From the Center of the Indenter To the Inner Wall | 110mm |
| Power Supply | AC220V±5%,50~60Hz |
| Executive Standard | ISO 6507,ASTM E384,JIS Z2244,GB/T 4340.2 |
| Dimensions | 405*290*480mm |
| Net Weight | 40KG |
WORK SHOP